01
HSIN6000A Fully Automatic Integrated Pressure Calibration Bench
Description
HSIN6000A fully automatic integrated pressure calibration table (comprehensive type) adopts standardized design, modular integration and integrated design layout, integrating pressure making, testing, acquisition and processing. Multi range and multi-media automatic pressure generation and regulation. The pressure making part has five optional units: vacuum, micro pressure, air pressure, hydraulic pressure and water pressure. Each measuring range is completely independent, which can not only play a function in combination, but also be used flexibly alone. It is widely used to calibrate various pressure gauges, pressure transmitters, pressure switches, pressure sensors and other industrial process instruments.


Features:
1.Full automatic verification, automatic and manual free switching;
2. Simple operation and convenient maintenance;
3.The software has powerful functions, icon management menu and automatic leak detection function, and supports information management, calibration parameter setting, execution and data analysis;
4.Professional development for customer needs, customers can customize according to their own needs;
5.The pressure control is instantly stable, leaving the tedious manual operation of servo voltage regulation;
6.With the development of three-dimensional work in the whole range, the degree of integration is higher and stronger;
7.Each unit is independent and highly targeted, which can be matched and combined flexibly;
8.Each output port has a built-in 50 micron filter to improve the service life;
9.The rising and falling pressure approaches point by point, which fully conforms to the verification regulation;
10.The system has many protection functions such as overheating and overload.
Pressure Unit
| Name | Pressure range | Automatic voltage stabilizing time of verification point | Pressure stabilit | Working medium |
| Vacuum unit | (-0.095~0)MPa | ≤15 seconds | 0.01%F.S/s | air |
| Micro pressure unit | (-10~10)kPa | ≤15 seconds | 0.01%F.S/s | air |
| Air pressure unit | (0~4)MPa | ≤15 seconds | 0.01%F.S/s | air |
| Oil pressure unit | (0~60)MPa | ≤25 seconds | 0.01%F.S/s | transformer oil |
| Water pressure unit | (0~60)MPa | ≤25 seconds | 0.01%F.S/s | purified water |
Electrical measurement technical index
Resolution
0.1μA
0.1mA
/
/
| Electrical measurement function | Range | Accuracy |
| Current measurement | (-30~30)mA | 0.02%F.S+0.003%RD |
| Voltage measurement | (-30~30)V | 0.02%F.S+0.003%RD |
| DC output | DC24V | ±0.5V |
| Pressure switch measurement | On off | Less than 12 ms |

Application
Environment condition
Ambient temperature: (- 10 ~ 50) ℃
Relative humidity: < 95%
Atmospheric pressure: (86 ~ 106) kPa
Storage temperature: (- 20 ~ 70) ℃
Online software
1.Comply with national metrological verification regulations
①JJG 52-2013Verification regulation of spring element general pressure gauge, pressure vacuum gauge and vacuum gauge
②JJG 49-2013 Verification regulation of spring element precision pressure gauge and vacuum gauge
③JJG 882-2019 Verification regulation of pressure transmitter
④JJG 875-2019 Verification regulation of digital pressure gauge
⑤JJG 544-2011Verification regulation of pressure controller
⑥JJG 860-1994 Verification regulation of pressure sensor (static)
⑦JJG 270-2008 The verification regulation of hemopiezometer
2.It runs in Windows2000, XP, win7, win10 and other operating systems with powerful functions, which is convenient for operation, collection of verification data, automatic generation of verification records, automatic generation of verification certificates, automatic printing, measurement management, etc
3.It has ledger database management function.